Search found 98 matches

by Luc
Tue Feb 20, 2024 6:41 am
Forum: OpenRAMAN
Topic: Alignment Question
Replies: 9
Views: 436

Re: Alignment Question

Which camera model of the ProSilica are you using exactly? Exposure time is typically around ~10 seconds on the Starter Edition for precision measurement to fully expose the sensor and get the highest dynamic range but it's still possible to recognize spectra with lower exposure time, down to 100 ms...
by Luc
Mon Feb 19, 2024 10:22 am
Forum: OpenRAMAN
Topic: Alignment Question
Replies: 9
Views: 436

Re: Alignment Question

Hi, It is possible to use the camera software directly for the alignment. At which stage of the alignment process are you stuck ? Also, but that's more of a general remark, it is not recommended to use a color camera sensor because you are throwing away more than 75% of the signal when using a green...
by Luc
Sun Feb 11, 2024 12:04 pm
Forum: OpenRAMAN
Topic: Converting CCD Image to Raman Spectrum
Replies: 1
Views: 243

Re: Converting CCD Image to Raman Spectrum

I don't have any names on the top of my head but there are surely some. At the very least, you can use the Matlab code that I gave here : http://www.open-raman.org/build/breadboard/assembly/ The process of converting an image to a spectra is not very complicated and can be split into two steps: (1) ...
by Luc
Thu Feb 01, 2024 7:00 am
Forum: OpenRAMAN
Topic: Open Raman Software Issue: Camera Not Detected
Replies: 2
Views: 636

Re: Open Raman Software Issue: Camera Not Detected

Hello, It's possible to add support for other camera brands but you need to make a program to "bridge" the functions from the SpectrumAnalyzer software to the camera own functions. I did it for Daheng cameras and it's a couple of days of work when you know what you're doing. It however req...
by Luc
Mon Jan 22, 2024 10:40 am
Forum: OpenRAMAN
Topic: Beginner in spectroscopy
Replies: 1
Views: 304

Re: Beginner in spectroscopy

Hello rama, That's a lot of questions -- I will try to answer them one by one. No questions are stupid btw and we're all here to help :) Generally speaking, professional Raman spectrometers are indeed expensive -- reason for which I started the OpenRAMAN project. On the other hand, you should not ne...
by Luc
Tue Nov 07, 2023 3:06 pm
Forum: General
Topic: Some questions about EC meter
Replies: 2
Views: 3397

Re: Some questions about EC meter

The idea of using a reference value was to compensate for small drifts in the amplitude of the applied sine wave.

I haven't test without it but when I designed the circuit I wanted to be safe and included as baseline.
by Luc
Mon Oct 09, 2023 4:19 pm
Forum: Autocollimator
Topic: Parts list
Replies: 1
Views: 3795

Re: Parts list

The people from Strathclyde were kind enough to make a text version of the list at the bottom of the page here:

https://focalplane.biologists.com/2022/ ... f-systems/
by Luc
Thu Aug 03, 2023 1:50 pm
Forum: OpenRAMAN
Topic: Raman Thermometer
Replies: 13
Views: 10378

Re: Raman Thermometer

The generic background correction method is not ideal for quantitative work.

The easiest is to export the raw spectrum in csv and, using Matlab or excel, remove a slanted line at the base of the peak. It works very well for isolated sharp peaks but it's more tricky for broader peaks.
by Luc
Sat Jul 22, 2023 7:03 am
Forum: OpenRAMAN
Topic: Signal to noise calculation
Replies: 2
Views: 1847

Re: Signal to noise calculation

There are two variables that can be used to define the noise of a spectrograph which are often mixed up. Here is a bit of clarity in that mix: Signal to Noise Ratio (SNR) which is technically how much noise is contained in every pixel of your data. To obtain it you typically take at least 10 spectra...
by Luc
Fri Jul 21, 2023 3:51 pm
Forum: OpenRAMAN
Topic: Raman Probe design WIP
Replies: 5
Views: 4990

Re: Raman Probe design WIP

The issue with single mode fiber is that it's very difficult to couple the signal into it. And the big disadvantage of the larger core fiber is that you loose a lot of light. I remember that my boss tried to make an OpenRAMAN reproduction at the office except he decided to change the design to use a...